Mouse Embryo Assay for Assisted Reproduction Technology Devices January 2021

Final
Docket Number:
FDA-2019-D-2105
Issued by:
Guidance Issuing Office
Center for Devices and Radiological Health

FDA is issuing this guidance to outline recommendations on conducting the mouse embryo assay (MEA) to support premarket submissions and lot release of assisted reproduction technology (ART) devices. The majority of ART devices directly or indirectly contact gametes (i.e., sperm and oocytes) and/or embryos during their intended use. MEA is used to assess the potential for embryotoxicity of devices that contact gametes and/or embryos. Several classification regulations under 21 CFR part 884 include special controls that require MEA testing or information. MEA may also be used to support premarket submissions for other devices that are intended to contact gametes and/or embryos during their use. However, there are no voluntary consensus standards that describe how to conduct the MEA. This guidance provides recommendations for conducting the MEA to support premarket submissions for devices that are intended to contact gametes and/or embryos and to comply with the special controls for those devices classified under 21 CFR 884 that require MEA testing or information.


Submit Comments

You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))

If unable to submit comments online, please mail written comments to:

Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852

All written comments should be identified with this document's docket number: FDA-2019-D-2105.

 
Back to Top